6.1b Mounting the TMCF in a 180
Use the following steps to properly mount and install the transducers in a 180
1. Familiarize yourself with the components of the TMCF as shown below.
2. Prepare the pipe where you intend to place the TMCF by making sure it is clean and free of loose material.
Sanding, though usually not required, may be necessary to take off any high spots. Be careful to preserve the
original curvature of the pipe.
3. If not already connected, connect the chain from one block to the screw hooks on the other block. The distance
between the blocks depends on where you are placing them on the pipe. You only need to estimate the distance. If
necessary, the distance can be adjusted later.