3.6 Base3DPM Status Indicator LEDs

Mezzanine Modules
A mezzanine module conditions the signal from a non-conductivity detector. The mezzanine
plugs into the Base3 DPM in order to tailor the DPM for a specific measurement.
Three primary types of mezzanine are available to accommodate FID and FPD detectors, and
various detectors that produce a scaled analog output (AI) mezzanine. Some of the
mezzanines have a dual range function for maximum flexibility. See the table below for details
relating to the various mezzanine options.
The AI mezzanine can be used for reading a detector voltage signal from a specialized or third
party detector, such as the Valco PDD, where the device only supplies a scaled voltage output.
The AI signal will be treated like a normal detector signal, with a 50% balance range.
Table 3-1
Mezzanine
2020960-001
2020960-003
2021328-002
2021328-001
2021328-003
2021326-001
1901614-001
Dummy Plug
3.6
Base3DPM Status Indicator LEDs
Figure 3-11
28
Mezzanine Part Number Descriptions
Detector Sub
Usage
Module Type
FID
Low level FID
Standard FID
Large Scale FID
FPD
FPD
FPD, 0.18 Hz Filter
Universal
Voltage AI
When Base DPM is Filament only,
and no mezzanine required
Heater 1
Power Active
Temp Limit
Overtemp
Alternate Range
Base3DPM Status LEDs
Heater 2
Power Active
Temp Limit
Overtemp
Flameout
Ignite
Normal
Fault
Warning
Service Manual, May 2018, A5E42019847001
Normal
Alternate
Range
Range
0.2nA
none
1nA
20nA
100nA
1000nA
100nA
none
100nA
none
±1V
±10V
none
none
Maxum Edition II Detectors