HP 3478A Technical Manual page 57

Dilgital multimeter
Hide thumbs Also See for 3478A:
Table of Contents

Advertisement

Hewlett-Packard Model 3478A
Digital Multimeter
Serial Number_____________
AC Volts Test
Step#
Input to
3478A
1
Open
2
Open
3
.03V,20KHz
4
0.3V,20KHz
5
0.3V,20KHz
6
1.5V,20KHz
7
3V,20KHz
8
3V,20KHz
9
30V,20KHz
10
30V,20KHz
11
300V,20KHz
12
0.3V,50KHz
13
3V,50KHz
14
30V,50KHz
15
300V,50KHz
16
0.3V,100KHz
17
0.3V,100KHz
18
3V, 100KHz
19
15V,100KHz
20
30V,100KHz
21
300V,100KHz
22
30V,300KHz
23
3V,50Hz
24
3V,20Hz
25
3V,20Hz
PERFORMANCE TEST CARD
1 YEAR LIMITS
Set-Up and
Configuration
Press TEST/RESET
ACV Function
300mV Range
030.250mV
300mV Range
301.033mV
3V Range
0.30180V
3V Range
1.50492V
3V Range
3.00882V
30V Range
03.0180V
30V Range
30.0882V
300V Range
030.201V
300V Range
301.092V
300mV Range
301.927V
3V Range
3.01410V
30V Range
30.1290V
300V Range
301.860V
300mV Range
306.102mV
3V Range
0.31140V
3V Range
3.03975V
30V Range
15.2085V
30V Range
30.3345V
300V Range
304.605V
30V Range
33.4020V
3V Range
3.01483V
3V Range
3.03522V
4 Digit Disp
3.0776V
Test Performed By______________________
Date_____________
Reference Temperature__________________
High
Reading
Limit
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
5
TM 11-6625-3071-14
Low
Test
Limit
Pass
029.750mV
_____
298.967mV
_____
0.29820V
_____
1.49508V
_____
2.99118V
_____
02.9820V
_____
29.9118V
_____
029.799V
_____
298.908V
_____
298.073V
_____
2.98590V
_____
29.8710V
_____
298.140V
_____
293.898mV
_____
0.28860V
_____
2.96025V
_____
14.7915V
_____
29.6655V
_____
295.395V
_____
26.5980V
_____
2.98517V
_____
2.96478V
_____
2.9224V
_____
Test
Fail
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____

Hide quick links: