Canon R-F10 Operation Manual page 29

Full auto ref
Table of Contents

Advertisement

6
Move measurement head toward the other eye
If "
R–L MEASURE
head automatically moves to the other eye.
If the setting is "
7
Alignment and measurement of the other eye are performed
Alignment and measurement are
performed automatically in the same
way as measuring the right eye is
done when the pupil of the left eye
has been detected.
After both eyes are measured, pupil
distance (PD) will be displayed.
8
The result is printed out
If "
" and "
PRINT
result will be printed out automatically after both eyes are measured.
If the setting is "
– Result will disappear from the monitor. However, it is saved in memory and can be
referred to on the DISPLAY mode screen and/or can be printed out again until the
next measurement is performed.
– If "
" is set to "
TRANS
Measurement data will be transferred to external instrument.
– If "
COUNT
" is set to "
Serial number will count up after data is printed/transferred.
NOTES: 1. Results will be deleted from memory at the measurement after printing/
transfer is performed.
2. It is recommended that a hard copy of the printout be made if you wish to
store it for a long time, because printouts on thermal paper are apt to
deteriorate.
4.2 Measurement in FULL AUTO Mode
" is set to "
" in SET mode (→ section 6.1), the measurement
ON
", press the
switch.
OFF
R/L
" are set to "
AUTO PRINT
", press
switch.
OFF
PRINT
"(→ section 6.2):
ON
ON
"(→ section 6.1):
– 17 –
PD: 67
VD:13.5
FULL AUTO(3)
LEFT
SPH
CYL
1:
-0.62
0.00
2:
-0.75
0.00
3:
-0.75
0.00
[
-0.75
0.00
" in SET mode (→ section 6.1), the
ON
No. 00001
AX
180
180
180
180
]