en | Programming
Enter easy unit ID programming
Remote programming (incoming)
The device acknowledges by announcing "Setup activated" and
enters the incoming remote programming mode.
6.3

Test mode

You can set the Carephone 62 into test mode. Pressing the
wireless transmitter or the emergency call button generates an
acoustical signal without starting an alarm.
Enter test mode
During 3 minutes, the red lamp flashes every two seconds and
the Carephone 62 emits a short beep every five seconds.
NOTICE!
Every time the Carephone 62 receives a signal from a registered
wireless transmitter, the three-minute time frame is reset.
If no signal from a registered wireless transmitter is received,
the Carephone 62 returns to standby.
F.01U.141.162 | V3 | 2012.11
Press P, 3 and 0 together for at least two seconds.
The unit goes directly to Programming Step 30.
Type the unit ID (device number) with numeric keys, the
unit ID can be up to 12 digits long.
Press C to confirm.
Press buttons <Yellow> <Green> <Grey> together for at
least two seconds.
Press P and A together for at least one second.
The Carephone 62 announces "test".
Press 9, 0 and C.
Press the button of a registered wireless transmitter.
When the Carephone 62 receives the signal from a
registered wireless transmitter, it emits a long beep.