HP 8481A Operating And Service Manual page 30

Power sensor
Hide thumbs Also See for 8481A:
Table of Contents

Advertisement

HP 8481A/82A/83A
08481-90049
6. Using tweezers, position the ends of the gold wires over the
electrical pads.
DO NOT tighten clamp screw 6 excessively or the FET circuit may be
broken.
Caution
«
Place and hold plastic clamp 16 over the gold wires. As you tighten
the clamp screw, watch the compression spring. Tighten the clamp
screw 7 only until the spring coils touch. Any further tightening
could damage the FET circuit.
Place the upper chassis in position and insert cap screws 1, 2, 10,
and 13.
Tighten 1, 2, 3, and 4.
Tighten 10, 11, 12, and 13.
Place the plastic shells, magnetic shields, and the chassis together
as shown in figure 1. Snap the plcistic shells together.
7.
8.
9.
10.
11.