Philips 4523 001 21821 Manual page 12

Wireless foot switch
Table of Contents

Advertisement

3 Operation
2 Connect the charger to the charging port. Ensure that you align the red dot on the charger connector
with the marker on the charging port.
The battery indicator on the wireless foot switch flashes green while the foot switch is charging.
NOTE
If you use the wireless foot switch with the charger connected, ensure that the wireless foot switch
is positioned to allow the charger to be easily disconnected from the foot switch if necessary. You
may need to disconnect the charger to restart the wireless foot switch. If care is not taken when
disconnecting the charger, damage may be caused to the charger or the wireless foot switch.
3 To stop charging the wireless foot switch, disconnect the charger from the charging port by doing the
following:
Figure 4 Disconnecting the charger cable
a Hold the connector firmly and push it gently toward the wireless foot switch, and hold the
connector in that position.
b Using your other hand, grip the collar and pull the connector out of the charging port.
c Put the cap back on the charging port on the back of the wireless foot switch.
NOTE
If you disconnect the charger while X-ray is being performed (while a pedal is being pressed), image
acquisition stops and should be restarted, if needed, by releasing and pressing the pedal again.
A complete charge cycle takes more than 12 hours. Ensure that the wireless foot switch battery is charged
at least every week. We recommend that you charge the battery at the end of every day, or when the
battery indicator turns red. The battery has built-in safety devices to protect it from overcharging.
The duration of use after charging depends on the model of your foot switch. For more information
about the model types, see
• For the type 1 wireless foot switch, a complete charge lasts for one week of use.
• For the type 2 wireless foot switch, a charge of between 6 to 8 hours provides up to 8 hours of
continuous use.
NOTE
If the battery of the wireless foot switch is depleted within 2 days after a complete charge, contact
technical support for a replacement battery. The battery may only be removed and replaced by a
qualified service engineer.
Storing the wireless foot switch for longer than 12 months without recharging the battery may damage
the battery.
Storing the wireless foot switch for longer than 24 months without recharging the battery will cause
battery failure.
Wireless Foot Switch Addendum
Equipment Labels (page
12
Charging the Wireless Foot Switch Battery
15).