The syntax is
Some BASIC examples:
10
CLEAR 7
to clear all HP-IB instruments, or
10
CLEAR 719
to clear an addressed instrument.
Related statements used by some computers:
RESET
CONTROL
SEND
Trigger
Trigger initiates a single event, such as a single sweep or an instantaneous measurement, from all
instruments or an individually addressed instrument. If the HP 8340B/41B is in analog-sweep mode
TRIGGER starts the sweep; if in CW, manual sweep, or fast phaselock (code FP) mode TRIGGER
increments the frequency. When the upper frequency limit is reached (either the stop frequency for
swept modes, or maximum instrument operating frequency for CW mode) the next TRIGGER com¬
mand will initiate a retrace to the starting frequency. The syntax is
A BASIC example:
10
TRIGGER 7
to simultaneously trigger all HP-IB instruments, or
100
TRIGGER 719
to trigger an instrument at address 19, or
100
TRIGGER 719, 721 , 712
to trigger instruments at addresses 19, 21, and 12.
Related statements used by some computers:
RESUME
SEND
The preceding statements are primarily management commands that do not incorporate instrument-
specific programming codes. The following two statements do incorporate programming codes, and
are used for data communication.