HP 3476A Operating And Service Manual page 22

Table of Contents

Advertisement

Section V
Table 5-1. Required Test Equipment.
Instrument Type
Characteristics
Recommended Model
Digital Multimeter
DC Volts: 1 V, 10 V, 100 V
Accuracy: . 05%
Input Resistance: .>-.-
to MO
AC Volts: . 1 V, 1 V ranges
Accuracy: .5%
Input Resistance: 10
MO
-hp- 3465A
DC Standard
Output: . 1 mV to 1000 V
Accuracy: .02%
-hp- 7408
AC Calibrator/High
Voltage Amplifier
Frequency: 45 Hz to 10 kHz
Output: 10 mV to 1000 V
Accuracy: 0.1%
-hp 745A/746A
Meter Calibrator
Output: 1 A
Accuracy: 0.1%
-hp- 6920B
Electronic Counter
Frequency: 10 kHz
Accuracy: 0.01%
-hp- 5300A/5302A
Power Supply
Output: 5 V, 1 A
-hp- 6294A
Resistive Decade Box
Ranges:
to 0, too sz, 1 ks
-
2,
10 kn, 100 4(12 and 1
ros
-
2
Steps
Accuracy: .05%
General Radio Model
GR 1433Z
Resistors
10 M12 ± 0.1%
1 MO ±
0.1%
300 krt ± . 1%
1 kit ± . 1%
10 K ± . 1%
-hp- Part No. 0698-8194
-hp- Part No. 0698-6369
-hp- Part No. 0698-6332
-hp- Part No. 0698-3491
-hp- Part No. 0698-4157
Model 3476A