Performance Tests
4-36
h. Repeat Steps e thru g until time interval measurements have been taken for each
voltage listed in Step d. The repeating sequence is:
1. Set the time interval to 200 ns.
2. Increment the dc voltage by 1 V.
3. Record the new time interval on the work sheet.
i. Each value entered in column ''T"'' is the incremental time interval for a 1 volt input
step, plus 200 ns. The accumulative time interval is the sum of all the preceeding incremental
time intervals, less 200 ns per time interval. Find the accumulative time intervals (yn) and
enter the results in column ''C'', and in the ''Measured Results'? column on the Perfor-
mance Test Record. If, for example, all the incremental time intervals are 700 ns, the ac-
cumulative time intervals would be:
0, 500 ns, 1000 ns, 1500 ns, . . . 5000 ns
j. Total all the entries in column ''C''.
1. Total all the entries in column ''D''.
Model 3336A/B/C