Field Installation Of Options - HP 5315 A Operating And Service Manual

100 mhz universal counter
Table of Contents

Advertisement

Model 5315A/B
Installation
2-29.
FIELD INSTALLATION OF OPTIONS
2-30.
The fo
ll
owing tables provide inst ructiona
l
procedures
fo
r
field installat ion of Options
001
and
002.
Table 2-1 Option
001
In51;)lIal;on
OPTION
001,
A7
TCXO
ASSEMBLY
Parts
requ ired
:
(11
A7 TCXO
Assembly,
OS317-60007
(2)
%2 X ¥S
Pan HD
Pozi
Screw, 236O--{)197
(2)
%2
x V.
Hex
Nut,
2500-0001
(2)
No.6
l
ockwasher,
2190-0008
Non
Installalion of Options
should
be performed by
qualified service
personnel
only.
1.
Remove the AC power
cord.
2.
Remove the
instrument top
cover. For Model
5315A,
Ihis requires removal
of
the
four
screws
accessible
from
cabinet
bottom.
For Model
5315B, loosen
the single screw
on the
top
center
of
t
he
rear
panel.
3.
Insert
the
A7 TCXO assembly,
wit
h components
fa
cing
to
the outside of the instrument, inlo
11
of
the
A1 Motherboard assembly.
4.
Secure
the
assembly by insta
lli
ng
the two
pozidrive
screws,
through
l-brackets
on A7,
to
the
Motherboard.
5.
Slide
the
motherboard
INT/
E
XT switch 53,
to the EXT pOSition (toward rear
of inSifumenl
).
6.
Replace
the top cover.
This completes the
installation of Option
001.
Table 2-2 Option 002
Installation
OPTION 002, BATIERY PACK WITH A6
CHARGER
ASSEMBLY
Parts
required
:
(1
)
Battery Pack Retrofit
Kit,
05315--60105
NOTE
Option 002 may be instal led only
in
Mode
l
531SA. Installation shou ld be
performed by qua
li
fied
service
personnel only.
1.
Remove the
AC
power
cord.
2.
Remove
the instrume
nt
top cover
by
removing
the
four
screws
accessible from the cabinet
bottom.
3.
Position
the
bracket assembly,
with
A6 and
BTl
insta lled, onto
the
six mounting posts on
the
inside top
cover,
placing
the
charger
toward the front and the battery toward the rear
of
the
instrument. Secure with
six
press-on retainers supplied.
4.
Connect
J1
of the
A6
Cha rger assemby 14 to the
Al Motherboard
assembly
using
cable
harness
supplied
NOTE
Connectors on
this
harness are polarized; do
not
force
the
connector.
5.
Replace
the top
cover.
Th
is
completes the installation of Option 002.