Setting the internal sweep
— pushbutton STD BY / CONT must be set to STD BY
— adjust the sweep period with the knob PERIOD
Starting a single sweep is done by pressing the pushbutton TRIG
once, the continuous
sweep is
realized by setting the pushbutton STD BY / CONT
to CONT.
In addition to this a sweep action
can
be started by a triggering pulse at the SWEEP
TRIG
INPUT socket. Starting criterion is the
negative edge e.g. of a TTL signal.
The SWEEP
OUT socket at the rear side provides the instantaneous sweep voltage corresponding
to a distinct signal frequency for controfling an oscilloscope or x - y plotter. The voltage to frequen-
cy relationship is linear, a sweep-voltage of O V represents f+, 7. 4 sweep voltage of +5 V fstor:
Further control elements for manual controlling are the HOLD pushbutton to stop the sweep imme-
diately and RESET for resetting to fstant during the sweep.
Note:
Before executing an internal sweep, disconnect all cables from the SWEEP INPUT socket.
This input is not switched off during an internal sweep.
Utric
SWEEP
TRIG
INPUT
t
f
'stop
ere
fstop™ dial reading
.
muiltiptied by the
factor of the frequency
eis
range selector
9
fstaRT 7 adiustment of
FREQUENCY
Hz
ht
knob
START
START FREQ
sweep
range selector SWEEP
PERIOD
and knob PERIOD
time
SWEEP
OUT
Internal Sweep
Diagram of a sweep period with the sweep
output-voltage SWEEP OUT and the
triggering signal SWEEP
TRIG
INPUT.

Advertisement

loading