Page 2
Indications for Use/Intended Use If a fault should occur during operation of the unit, operate the disconnect device to switch the power The LCD Monitor is intended to provide 4K 2D color supply off, or disconnect the power plug. video displays of surgical and other medical images.
Page 3
IEC 60601-1 standards. If You are cautioned that any changes or in doubt, consult qualified Sony service modifications not expressly approved in this personnel. manual could void your authority to operate this equipment.
Page 4
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550MD. Otherwise, degradation of the performance of this equipment could result. If the LMD-XH550MD will be used adjacent to or stacked with other equipment, normal operation of the LMD-XH550MD under such configurations should be verified via observation.
Page 5
(50/60 Hz) typical commercial or hospital environment. magnetic field IEC 61000-4-8 Proximity 65 A/m 65 A/m Proximity magnetic fields of the LMD-XH550MD magnetic fields 134.2 kHz 134.2 kHz should be at levels characteristic of a typical Pulse modulation Pulse location in a typical commercial or hospital IEC 61000-4-39 2.1 kHz...
Page 6
Guidance and manufacturers declaration electromagnetic immunity The LMD-XH550MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-XH550MD should assure that it is used in such an environment. IEC 60601 Compliance Immunity test Electromagnetic environment ...
Page 7
If the measured field strength in the location in which the LMD-XH550MD is used exceeds the applicable RF compliance level above, the LMD-XH550MD should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the LMD- XH550MD.
Page 8
The LMD-XH550MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550MD. Otherwise, degradation of the performance of this equipment could result.
Page 9
Top side: 30 cm (11 in.) or more Warning on power connections for medical use Consult with Sony qualified personnel for the Customers in the U.S.A. and Canada should use following types of installation location. the following type of power cord.
Page 10
Firmly grip the bottom of this unit as shown www.sony.com/psa/warranty for important below. information and complete terms and conditions of Sonys limited warranty applicable to this product. For the customers in Canada SONY LIMITED WARRANTY - Please visit www.sony.com/psa/warranty for important information and complete terms and conditions of Sonys limited warranty applicable to this...
Preset menu .........33 Table of Contents Troubleshooting ..........33 Error Messages ..........34 Specifications ..........34 Precaution ............12 Dimensions ............40 On Safety ...........12 Licenses ............41 On Installation ...........12 Safety precautions for using this unit ..12 Precautions when connecting this unit to medical equipment .........12 To prolong the life of the unit ....12 Cautions for RESPONSIBLE ORGANIZATION...
When connecting the unit to medical Precaution equipment, refer to “Precautions when connecting this unit to medical equipment”. On Safety Precautions when connecting this unit to medical equipment Operate the unit only with a power source as specified in the “Specifications” section. Before using this unit for medical purposes, be ...
update of equipment connected to the IT- addition, over a long period of use, because of NETWORK; and the physical characteristics of the liquid crystal upgrade of equipment connected to the IT- display, such “stuck” pixels may appear NETWORK.
On Security the screen continuously, or repeatedly over extended periods. SONY WILL NOT BE LIABLE FOR DAMAGES OF ANY KIND RESULTING FROM A FAILURE TO Images that may cause burn-in...
This function sharpens images. You can select the mode best suited for the resolution of the equipment used. On Repacking 1) A.I.M.E. is a registered trademark of Sony Do not throw away the carton and packing Corporation. materials. They make an ideal container which to transport the unit.
Page 16
Flat surface for better maintenance The design allows the user to easily wipe liquids and gel off the LCD panel and control buttons – ensuring a high standard of disinfection and cleanliness.
Location and Function of Parts and Controls Front Panel Power indicator When you press PORT A or PORT B when it is lit in white, an image from the input Indicator Operating State connector assigned to the PORT A/B is displayed.
Page 18
CUSTOM buttons Turns on or off the assigned function. You can adjust the assigned function by pressing the / buttons. (Refer to the Custom Button of the System Configuration menu on page 32 and of the Initial Setup menu on page 33.) The following functions are assigned in the default setting.
Rear Panel To use the connectors on the rear panel, remove the cable cover. For details on the cable cover, see page 23. Cable cover (L) cover to the right. The cable of the 12G-SDI 1 input/output connector can be connected or When connecting or disconnecting the cable of the connector on the rear panel, remove this disconnected without removing the cable cover...
Page 21
Active through output connector for serial digital component signals (4K/HD). SDI 1 and Note SDI 2 outputs are available. Consult with Sony qualified personnel at using 3G-SDI input/output connector (BNC type) this connector. input connector Caution Input connector for serial digital component For safety, do not connect the connector to signals (HD/SD).
Page 22
Be sure to connect equipment or cables 12V 2.5A (DC output) connector specified by Instructions for Use of this monitor Outputs 12 volts for connected external to the 12G-SDI/3G-SDI output connectors. If equipment. you connect unspecified equipment or cables, ...
Preparation Connecting AC input connector Make sure that the (On)/ (Standby) switch is set to (Standby). AC power cord Remove the cable cover (L). Slide down the cable cover (L) while sliding Plug holder for the the slide locks (2 pcs), which are located at supplied AC power cord the bottom of the monitor, in the direction of the arrow.
Attach the cable cover (L). Turning on the Monitor/Switching Input Slide the cable cover (L), while sliding the Settings slide locks (2 pcs), which are located at the bottom of the monitor, in the reverse Connect the plug of the AC power cord to an direction of step2.
Initial Setting If you have selected the wrong area, change the Color Temperature setting (page 28). When you turn on the unit for the first time after purchasing it, select the area and language To set the using language where you intend to use this unit. You can select one of seven languages (English, Chinese, Japanese, Italian, Spanish, German, and To set the using area...
Using the Menu The unit is equipped with an on-screen menu for making various adjustments and settings such as picture control, input setting, setting change, etc. Press the / button to select the tab. The selected tab is shown in blue, and setting items of the selected tab appear.
About the control navigation Adjustment Using the Menus Depending on the state, the operation buttons of the unit light as shown below: White light: Operable state. Green light: Operating state. Items Off: Unable to operate. The screen menu of this monitor consists of the following items.
Input Name Menu Setting AUX IN Setting Chroma Adjusts color intensity. The higher the Power Supply setting, the greater the intensity. The lower the setting, the lower the intensity. System Configuration (page 30) Brightness Adjusts brightness. Control Lock Contrast Adjusts contrast. OSD Setting A.I.M.E.
To display the corresponding signal with “Enhanced Format,” use a display setting for each input. Premium High-Speed HDMI cable within a length of 3 meters (Sony Menu Setting product recommended). 4K Scan Size Select the scan size for the display of the 4K equivalent signal from “Off,”...
Menu Setting Menu Setting Sets the sub screen position for the 2 Input Name Set the name of the 12G-SDI 1, 12G-SDI or 3 screens display (POP). 2, 3G-SDI, Display Port, HDMI, DVI-D input connector. Pattern Skip Sets the pattern that is skipped when ...
Page 31
Menu Setting Menu Setting Power On Setting Input Connector Name Power On Mode Select the setting when the monitor is turned on from the following settings. Select Port Control Lock Last: The setting when the monitor was last turned off. ...
Menu Setting Menu Setting Remote Backlight Control Off: Turns off backlight control. Remote Mode Sets the remote function when the Standard: REMOTE connector is connected to the Sets the backlight to standard optional Foot Switch FS-24. brightness mode. ...
Menu Setting Troubleshooting PORT A HD Select the clipping size of the HD Clipping Size/ signal from “Normal,” “4:3,” “5:4,” or “V PORT B HD Full” when displaying 2 or 3 screen Clipping Size displays. This section may help you isolate the cause of a Custom Button Assigns the function to the CUSTOM 1 problem and as a result, eliminate the need to...
Error Messages Specifications Picture performance When the following messages appear on the screen, turn off the power and contact an LCD panel TFT Active Matrix authorized Sony dealer. Pixel efficiency 99.99% Messages Description Viewing angle (panel specification) Fan Error A malfunction occurs in the fan.
Page 35
WARRANTY, OR FOR ANY OTHER REASON power cord (2) WHATSOEVER. Before Using This Unit (1) SONY WILL NOT BE LIABLE FOR CLAIMS OF CD-ROM (including the ANY KIND MADE BY USERS OF THIS UNIT OR Instructions for Use) (1) MADE BY THIRD PARTIES.
Page 36
Pin assignment 12V 2.5A (DC output) connector SERIAL REMOTE (RS-232C) connector D-sub 9-pin, female Pin number Signal Pin number Signal 12 V REMOTE (stereo mini jack) connector Pin number Signal On (short to sleeve) Off (open) 5V 2.0A (DC output) connector Pin number Signal...
Page 37
Available signal formats The unit is compatible with the signal systems shown below: Signal format 12G-SDI 1 12G-SDI 2 3G-SDI SDI Single Link SD-SDI 720×487/59.94i 4 : 2 : 2 YCbCr 10bit × × 720×576/50i 4 : 2 : 2 YCbCr 10bit ×...
Warning Dimensions Be sure to set the tightening torque value to the following value. Front Torque value: 1.7±0.1 N·m Make sure the tightening torque value is at this value. If the torque value is not appropriate, the mounting part may become damaged or the screws may become loosened, and in the worst case, it may cause injury or damage to the unit due to the unit falling.