To test bandwidth
To test bandwidth
This test checks the bandwidth of the oscilloscope.
Specification
Repetitive (all models)
dc to 2500 MHz (equivalent time)
Real Time
HP 54520A
dec to 2125 MHz
HP 54540A dc to 2125 MHz (3 or 4 CH on), de to 2250 MHz (1 or 2 GH on)
HP 54522A and 54542A
dc to 2500 MHz
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Signal Generator
1—500 MHz at ~200 mVrms
HP 8656B
Power Meter/Sensor
1-—-500 MHz +3% accuracy
HP 436A/8482A
Power Splitter
outputs differ by <0.15 dB
HP 11667A,
Cable
Type N (m) 24 inch
HP 11500B
Adapter
Type N (m) to BNC (m)
HP 1250-0082
Repetitive Test
Connect the equipment
a With the N cable, connect the signal generator to the power splitter input. Connect
the power sensor to one output of the power splitter.
b With an N-to-BNC adapter, connect the other splitter output to the channel 1 input.
Press
Recall
,thenpress
clr
tosetthe oscilloscope to default conditions.
Set the following parameters.
Menu (hardkey)
Function/Parameter
Setting
Vertical {channel keys)
(sensitivity)
100 mV/div
(all channels)
(input R)
50Q pc
Horizontal
(time/div)
200 ns/div
(Setup)
(mode)
repetitive
Display
(mode)
avg
# of avg
32
(graticule}
grid
Entry
Fine
(on)
Set the signal generator for 1 MHz at +5.0 dBm. The signal on the screen should be
about two cycles at six divisions amplitude.
Note: Do not exceed 6 divisions when making the bandwidth check.
Initiate a V amptd measurement.
,
Press
(blue) , V amptd ,and
1
(channel number, entry menu).
After the measurement settles (averaging complete, about 10 seconds) note the
Vamp (1) reading (bottom of screen): Vimuz =
mv.
Set the power meter Cal Factor % to 1 MHz value from the cal chart on the probe.
Then press dB[REF] to set a 0 dB reference.
3-9

Advertisement