Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
Protective Conductor Test ..........17 3 _______ Changes to previous version _____________________________________ 18 Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 3 of 18 Medical Solutions 08.05...
If an uninterruptible power supply (USV) is installed in the system, the voltage output of the USV must also be deenergized or the voltage output plug must be discon- nected. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 4 of 18 08.05...
Page 5
Avoid all contact with blood or other bodily fluids! Strictly observe the safety information in ARTD-002.731.37 regarding prophylactic measures against infectious diseases during service. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 5 of 18 Medical Solutions 08.05 CS PS 24...
(3/Fig. 1 / p. 8) und den OTV (4/Fig. 1 / p. 8) (optical isolation amplifier) in the imaging system container of the UROSKOP Access. - No OTV is installed if an endoscopic interface is present. Fig. 1: Imaging system container Pos.
9. Use the cable ties to secure the cables. 10. Connect the BNC cable located at the lifting base in the back wall of the device (moni- tor switchover URO). 11. Install the urodynamic interface in the back wall of the UROSKOP Access lifting base. Siemens AG SPL5-330.814.01.05.02...
ASPIA imaging system. • Press the Menu key until Adjust: Screen size appears. • Press the + key. - The sub-menu is displayed. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 12 of 18 08.05 CS PS 24 Medical Solutions...
• To save all settings, press the Menu key and the + key at the same time.. • Switch the UROSKOP Access off and back on again. Cutting off the menu bar (FLUOROSPOT Compact) NOTE This subsection refers exclusively to UROSKOP-Access systems with the FLC imaging system.
Page 14
3. For the scan converter model TVone CS-500A, this parameter is no longer selectable. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 14 of 18 08.05...
2. The parameters in the menu “Manual set“ (VGA left, VGA width, VGA top/4, VGA bott/4) can vary from customer to customer, depending on the factory-set monitor settings implemented by the manufacturer in the urodynamic system. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 15 of 18 Medical Solutions 08.05 CS PS 24...
Urodynamic interface Faulty screen display • To revert to the factory settings, switch off the scan converter, press and hold the + and - keys simultaneously, and restart the scan converter. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 16 of 18 08.05...
A sudden or unexpected increase in the measured values may in- dicate a defect in the ground wire connections (ground wire or contacts) - even if the limit value of 0.2 ohms is not exceeded. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 17 of 18 Medical Solutions 08.05 CS PS 24...
Page 18
“Cutting off the menu bar the scan converter“ (FLUOROSPOT Compact)“ new; “scan converter parameter table (FLUOROSPOT Compact imaging system)“ new Subsection “Final work “Protective conductor test“ new steps“ UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 18 of 18 08.05 CS PS 24 Medical Solutions...