Download Print this page

Sony AC-110MD Instructions For Use Manual page 18

Hide thumbs Also See for AC-110MD:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 14
Guidance and manufacturer's declaration – electromagnetic immunity
The AC-110MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
AC-110MD should assure that it is used in such an environment.
Immunity test
Conducted RF
3 Vrms
150 kHz to 80 MHz
IEC 61000-4-6
outside ISM bands
6 Vrms
150 kHz to 80 MHz
in ISM bands
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
6
GB
IEC 60601
Compliance level
test level
3 Vrms
c
6 Vrms
c
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the AC-110MD,
including cables, than the recommended separation
distance calculated from the equation appliance to the
frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
IEC 60601-1-2: 2007
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d = 2.0 √P 80 MHz to 2.7 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
b
range.
Interference may occur in the vicinity of equipment
marked with following symbol:
a
should

Advertisement

loading