en - Areas of application
38
For validation purposes please make sure that you follow the
loading instructions given on the template.
Before connecting hollow instruments, please make sure that all
hollow sections are free of blockages. Otherwise the cleaning solu-
tion will not be able to access all areas properly.
 Any hollow items that have become disconnected from their in-
jector nozzles, adapters or holders during processing must be re-
processed.
Before loading the machine and before starting a
programme
– Check that fittings such as irrigation sleeves and injector nozzles
are securely screwed into position.
 Make sure that all screw connectors are fitted with injector
nozzles, adapters, irrigation sleeves or blind stoppers to ensure
that all rinsing fittings in use are supplied with sufficient standard-
ised pressure.
Damaged fittings such as injector nozzles, adapters and irrigation
sleeves must not be used.
Rinsing positions not in use do not have to be fitted with blind
stoppers.
– Check that the mobile unit is correctly docked to the water path in
the chamber.