NOTE
The evaluation is not necessary in the case of a new determination.
File the report sheet in the system folder or log book.
Monitor 44cm, 100/120HZ
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding maintenance
procedures or safety checks.
A sudden or unexpected increase in the measured values may in-
dicate that a fault has occurred in the primary circuit of the power
supply (damaged insulation, damage caused by water ingress or
humidity, defective interference suppressor, etc.) - even if the limit
value of 2.5 mA is not exceeded.
SPR2-230.101.01.02.02
08.05
General information
Page 14 of 26