LG 14HK701G-W.AUB Owner's Manual page 10

Flat panel digital x-ray detector
Table of Contents

Advertisement

No.
LED Indicator
2
3
4
5
6
7
10
LED Color
Connecting the Control Box and the Detector A. This connector supply power (24 V
Ethernet image data.
Connecting the Control Box and the Detector B. This connector supply power (24 V
Ethernet image data.
Control Box supports 2 Detector connections. Usage is, one is for Bucky stand, the other is for table (bed). Generally, X-ray room of hospital installs 2
Detectors, Bucky stand and table type, it's far more convenient and efficient working environment. These 2 Detectors are not operated simultaneously,
control box selects the operating Detector by AWS command.
None
Connects AC power cord.
Ethernet port to transmit image/command between the Detector and PC.
This is to synchronize the Detector and X-ray Generator.
Control box power fuses are 4A, 250V to Type T fuse.
Power rating: T4L 250V
Status
2.1 A) to the Detector, transmits X-ray synchronization signals and
2.1 A) to the Detector, transmits X-ray synchronization signals and