Siemens SITRANS LG250 Operating Instructions Manual page 40

Guided wave radar
Table of Contents

Advertisement

Peak values, measure-
ment reliability
Peak values, additional
40
In another window you can carry out a reset of the two peak values
separately.
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in the menu item "Peak values,
measurement reliability".
The measurement can be influenced by the process conditions. In
this menu item, the measurement reliability of the level measure-
ment is displayed in mV. The higher the value, the more reliable the
measurement.
If you have selected interface measurement under the menu item
"Setup - Application", the peak values of the interface measurement
are displayed in addition to the peak values of the level measurement.
In another window you can carry out a reset of the two peak values
separately.
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad-
ditional".
This menu item displays the peak values of the electronics tempera-
ture as well as the dielectric constant.
In another window you can carry out a reset of the two peak values
separately.
SITRANS LG250 - Operating Instructions
PBD-51041053