How To Attach The Sterile Puncture Adapter (Ezu-Pa7V) - Hitachi C41V1 Instruction Manual

Table of Contents

Advertisement

8) After using the probe, clean, disinfect and if necessary
sterilize the probe immediately. If RVS is carried out,
clean and disinfect and if necessary sterilize the Magnetic
sensor attachment.
9) Store the probe and the Magnetic sensor attachment under
the conditions written in
"6. Maintenance and Safety Inspection" and "8.Specification".
・ Do not use the probe if the image and the frequency do not
match the probe. An incorrect acoustic output can result
in burns or other injuries to the patient.
・ Scan for the minimum length of time necessary for the
diagnosis and at the lowest possible output. There is the
possibility that the patient's tissues could be affected.
For details about the acoustic output, please refer to the
operation manual of the ultrasound diagnostic instrument.

3.2 How to attach the Sterile Puncture Adapter (EZU-PA7V)

How to attach the Sterile Puncture Adapter (EZU-PA7V) to the
probe is instructed below.
1)
Put the picks of the Sterile Puncture Adapter to the projection
on the probe. (See the top of Fig.6)
2)
Push the other end of the Sterile Puncture Adapter until the
dents of the Sterile Puncture Adapter are fixed at the grooves
of the probe. (See the bottom of Fig.6)
3)
Visually inspect the probe cover for any hole or tears.
Fig.6 how to attach the Adapter
CAUTION
Sterile Puncture Adapter
(EZU-PA7V)
Probe cover
- 8 -
Q1E-EP1432