Electrical Connections - ABB AZ200 770 Manual

Endura az20 oxygen monitor pumped reference air unit
Table of Contents

Advertisement

3.4 Accessing the terminal block
Referring to Fig. 3.3:
1. Loosen the 4 pump cover retaining screws
2. Remove the pump cover
C
3. Loosen cable gland
supply cable.
B
A
Fig. 3.3 Accessing the Terminal Block
A
.
B
.
to prepare for the incoming power
Pump Identification
Label
C

3.5 Electrical connections

3.5.1 Preparation
Referring to Fig. 3.4:
1. Prepare the incoming power supply cable as follows:
a. cut the outer sheath back to leave 100 mm (4.0 in.)
sleeved wires
b. cut the sleeved wires back to leave 5 mm (0.2 in.) tails
2. Pass the cable through cable gland
Endura AZ20 oxygen monitor | Pumped reference air unit | INF09/027 Rev. D
A
C
.
3 Installation
B
5