Kenwood FC-758 Instruction Manual page 24

Frequency counter
Table of Contents

Advertisement

The problems discussed above may be triggered by
standing waves.
In this case, the input circuit should be properly
terminated or the cable length needs to be adjusted.
The prescale input circuit is internally
w i t h a 5 0 - o h m resistance. T h i s means t h a t if the
signals output by a 5 0 - o h m external circuit are
measured, the problems possibly caused by standing
waves can be prevented.
In addition, unstable h i g h - f r e q u e n c y input signal
may cause an error.
In the period measurement mode, pseudo trigger
signal may cause measurement error. T h i s
measurement error can be reduced by increasing
the number of measurement cycles. T h a t is, a
longer gate t i m e value than the current one
should be employed for t h a t purpose.
A n y period
is a reciprocal of
frequency while any frequency is a reciprocal of
- its associated period.
If relatively l o w - f r e q u e n c y signal is measured, the
cycle measurement mode should be used instead of
the frequency measurement mode because higher
24
terminated
its associated
precision d a t a can be displayed.
If relatively h i g h - f r e q u e n c y signal is measured, the
frequency measurement mode should be used instead
of the period measurement mode because higher
precision measurement d a t a can be output.
E i t h e r measurement mode should be employed w i t h
a frequency of 10kHz in mind.
5 - 4 - 3 .
Measurement Delay
Display measurement data on Display ©
updated a t the moment when each measurement
period ( o r i n t e r v a l ) elapses. T h i s applies to every
measurement mode except for the event t o t a l
measurement mode. The G A T E L E D remains
active during the measurement period. T h i s period
is dependent on a selected measurement mode and
specified resolution value. T h a t is, the end of the
period can be defined differently.
significance in some cases. When you are about to
change resolution values, operational modes, or to
use the hold or reset function, you should take
t h a t delay into consideration because a new
measurement operation w i l l s t a r t .
is
T h i s delay has

Advertisement