Maintenance and service

11.7 Replacing option modules

Recommissioning the device
Note
Recommissioning the device
Recommission the device with the analyzer modules installed last. This guarantees that all
required data is restored on the new processing unit.
• No operating, user or factory data is saved on the basic device. The new PU is provided with
the most recent firmware.
• Restart the device with the analyzer modules installed last.
• Subsequent to switching on the device, the AM is recognized as new. Each AM retains its
parameter settings.
• The configuration of the basic device (operating data) is restored from the memory of
analyzer module AM1.
You can find additional information in section "Combined operation (Page 33)".
• The procedural descriptions in the operating manuals at "References (Page 135)" apply to
recommissioning.
• For further steps, please contact the Service (Page 149) department.
Note
Different IP address when replacing the processing module
Having different IP addresses after replacing the processing module can prevent network
communication with the device.
If you install a new processing module in the SIPROCESS GA700 device, the following IP address
is set as default in the factory: 192.168.0.2. Before you set up a network connection to the
device, follow the steps below:
• Check the IP address
• Set the IP address again prior to module replacement
You will find information on setting the IP address in the operating manuals (→ References
(Page 135)).
11.7
Replacing option modules
You can find information on replacing option modules in the section "Installing option modules
(Page 63)".
110
Rack-mounted device
Operating Instructions, A5E31874006-12