Solving General Problems
Problem
Transmission INOP
light on the receiver
is still lit when the
transmitter is
switched on.
Battery Low Light lit
on receiver.
INOP transmission
lamp is lit after the
patient has moved a
short distance away
form the receiver.
60
Possible Causes
Receiver and transmitter do not
have the same channel or serial
number.
Batteries in the transmitter are
exhausted.
Power in batteries is low.
Local Antenna:
Antenna not connected correctly.
Remote Antenna System:
Antenna cable not connected
correctly to receiver.
Solutions
Check channel number and the serial
numbers are the same on the receiver and
the transmitter.
Change the batteries in the transmitter
(refer to Chapter 3 for details.)
Change batteries.
Check antenna is connecting correctly.
Test the antenna system by bringing the
transmitter close to the receiver. If the
transmission is good, then the antenna
system is not functioning properly.
Contact your Philips Medical Response
Center.
Chapter 10 - Troubleshooting