Features
last temperature recall
Both the soother thermometer and the digital flexible-tip thermometer
have a memory function. When the soother thermometer and the digital
flexible-tip thermometer are switched on, the display automatically shows
the last measured temperature for approx. 2 seconds. The display then
shows the normal body temperature of 37°C.
Battery-low indication
Both the soother thermometer and the digital, flexible-tip thermometer
have a long-life battery which should last for about 2 to 3 years. If the
display shows a battery symbol, the battery is low. Replace the battery of
the flexible-tip thermometer (see chapter 'Replacing the battery').
Note: The battery of the soother thermometer cannot be replaced. If the
battery is empty, discard the soother thermometer.
Automatic shut-off
Both the soother thermometer and the digital flexible-tip thermometer
switch off automatically: - after 1 min if the temperature measured is
<32°C; - or after 10 min if the temperature measured is >32°C.
Error function
Both the soother thermometer and the digital flexible-tip thermometer
have an error function. In case of malfunction, the display shows an 'Err'.
Proceed as follows if this happens:
1
Press the on/off button I to switch off the thermometer.
2
Press the on/off button I to switch on the thermometer again.
Cleaning
Do not use abrasive cleaning agents, anti-bacterial cleaning agents or
chemical solvents. Excessive concentration of cleaning agents may
eventually cause plastic components to crack. Do not use cracked
parts.