Installing/mounting
4.4 Establishing the measuring range
4.3
Setting up the measuring points
①
②
Figure 4-2
4.4
Establishing the measuring range
Calculating the measuring range with process media with a density ≠ 1000 kg/m3 (process medium ≠
water)
p = ρ * g * h
with:
ρ = density of the process medium
g = local gravitational acceleration
h = maximum level
18
Cable box
Anchor clamp
Setting up the measuring points
③
Pressure transmitter
SITRANS LH300 Transmitter for Hydrostatic Level
Compact Operating Instructions, 01/2019, A5E37897183-AC