VA301D2/GasPoint II Calibration Procedure

IF YOUR TRANSMITTER WAS PURCHASED BEFORE APRIL 1, 2004, CONTACT TECHNICAL SUPPORT
LINE AT 1-800-563-2967
1. Warm-Up Period
Turn the unit on for a minimum of fifteen (15) minutes.
Check Calibration Information for specific warm-up time.
Gas
Combustible
Infrared Combustible 0 – 102% LEL
CL2
CO
ETO
H2S
HCL
HCN
HF
NO
NO2
O2
SO2
SIH4
F2
500 ml/min = 30 l/h on regulator
2. Connecting the Hardware
Plug the calibration adaptor onto the gas sensor inlet.
Screw the regulator to the calibration Span gas cylinder for the Span adjustment or Zero gas cylinder for
the Zero adjustment and set the appropriate flow rate.
Connect the regulator outlet to the calibration adaptor.
VA301D2 / GasPoint II
GasPoint II CALIBRATION INFORMATION
Scale
Warm-up
0 -102% LEL
15 min.
15 min.
0 - 15.3 ppm
15 min.
0 - 255 ppm
N/A
0 - 20.4 ppm
15 min.
0 - 51.0 ppm
15 min.
0 - 25.5 ppm
15 min.
0 - 51.0 ppm
15 min.
0 - 10.2 ppm
15 min.
0 - 102 ppm
15 min.
0 - 10.2 ppm
15 min.
0 -25.5%
N/A
0 -10.2 ppm
15 min.
0 -10.2 ppm
15 min.
0 – 2.6 ppm
15 min
Regulator
103 Liter
Span Calibration
58 Liter
Gas Cylinder
Span Calibration
Gas Cylinder
17 Liter
Span
Calibration
Gas Cylinder
17L, 58L or 103L Calibration Gas Cylinder
Calibration Handbook
Factor
Flow Rate
Table II
500 ml/min 1% Volume CH4 = 20% LEL CH4.
Table III
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
0.2
500 ml/min Nitrogen dioxide surrogate for HF
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
N/A
500 ml/min
1.3
500 ml/min CL2 surrogate gas for F2
GasPoint II
Calibration cap
Calibration tubing
15
Notes
2.5% Volume CH4 = 50% LEL
CH4.