7. Handling
Do not hold the FPI connector and sensor unite cable when lifting and/or moving the FPI.
Please be careful not to drop the FPI. If dropped, check appearance and do not use FPI if it
has abnormality.
8. Ingress of Spill Liquid
Do not spill a liquid (Blood, Body fluid, Cleaning liquid, etc) on the FPI. When spilled a liquid,
please wipe it off with soft cloth promptly, and clean the surface.
When a large quantity of blood or body fluid might attach, please use sanitary cover from a
sanitary point of view.
9. Ingress of Liquid
Do not splash the FPI with any liquid. Occasionally that is a cause of damage.
Wipe of once when it get wet.
10. Cleaning
When cleaning the surface of FPI, please wipe the surface by the soft cloth soaked with water,
oxygenated water or dilution ethanol. Do not use other medicines.
11. Load
Please do not apply excessive load on the FPI.
Maximum load: 150 kg @In the whole
100 kg @φ40 mm of the center
12. Unnecessary X-ray, Qualified Personnel
The FPI is used combined with X-ray tube assembly. Only Qualified personnel for X-ray handling
procedures can use the FPI.
For set-ups and services, only knowleged enigineers or special service personnels are able to
handle the FPI.
13. X-ray Shield
When using this device, unnecessary X-ray irradiation must be avoided.
(1) Please follow the "minimum distance between the X-ray tube focal point and X-ray incidence
plane" and "maximum tube voltage" specified in the technical data.
(2) This devise does not absorb 100% X-ray.
It requires X-ray protection measures for X-ray system to comply with the IEC 60601-1-3.
14. X-ray Specifications
Please follow the ratings below.
(1) Maximum tube voltage ........................................................................................... 150 kV
(2) Minimum tube voltage ··········································································· 40 kV
(3) Maximum X-ray incident dose rate ......................................... 8.73 x 10
(4) Minimum distance between the X-ray tube focal point and X-ray incidence plane .. 73 cm
-19-
FDX3543RP
-3
Gy/min [1R/min]