GE TONOPORT V Service Manual page 23

Hide thumbs Also See for TONOPORT V:
Table of Contents

Advertisement

Revision H
Firmware Loading Procedure for FW1.4: Loading Procedure
Switch off TONOPORT V.
Disconnect TONOPORT V from PC.
Check the correct Firmware version on TONOPORT V and set the correct
language:
Step
1
Switch on TONOPORT
V
2
Press "INFO" 5 times
3
Press "START"
4
Press "INFO" to select
the correct language
(which was displayed at
Step 1 of the Loading
Procedure)
5
Press "START" to save
the language setting
Screenshot of a successful loading procedure with Hyper Terminal:
TONOPORT V
2001589-023
Action
HH:MM (time)
H 5
14 : 01
(14 = Version FW 1.4
01 = German language)
14 : 01 (German)
14 : 02 (English)
14 : 03 (French)
14 : 04 (Italian)
14 : 05 (Spanish)
HH : MM (time
Diplay
7-5

Hide quick links:

Advertisement

Table of Contents
loading