Alarm Parameters - Panasonic MCO-170AICL Operating Instructions Manual

Co2 incubator

Advertisement

ALARM PARAMETERS

1. Press Menu key to lead the Menu screen.
2. Press Tools #1 key to lead the Tools #1 screen.
3. On the Tools #1 screen,
・Press Alarm Setting #1 key to lead the Alarm
Setting #1 screen, it is possible to set automatic set
temperature alarm and automatic set CO
density
2
alarm (go to procedure 4).
・Press Alarm Setting #2 key to lead the Alarm
Setting #2 screen, it is possible to set alarm delay,
door alarm delay, ring back and remote alarm (go
to procedure 5).
4. On the Alarm Setting #1 screen, input each
parameter. Press Apply key to save the input value.
The display returns to the Tools #1 screen.
●Each parameter setting
・Temp. Alarm:
When the chamber temperature exceeds the scope,
the set temperature ± the set value of Automatic set
temperature alarm, the alarm is activated. Settable
o
o
o
range: ±1.0
C~±5.0
C, factory setting: ±1.0
C。
・CO
Alarm:
2
When the chamber CO
density exceeds the scope, the set CO
density ± the set value of Automatic set
2
2
CO
density alarm, the alarm is activated. Settable range: ±0.5 %~±5.0 %, factory setting: ±1.0 %。