Hitachi EUP B715 Instruction Manual page 35

Biopsy probe
Table of Contents

Advertisement

B) Magnetic sensor attachment and spacer for EZU-RVS2
1) Before immersing the device, it is recommended to test the
concentration of disinfectant solution before each usage.
solution Cidex® OPA is ready for use and does not need to be diluted.
Test strips to verify that the appropriate concentration of Cidex®
OPA is correct are available by manufacturer. Test strips will
indicate a concentration above the Minimum Effective Concentration
(MEC). Please also note the expiration date of the test stripes.
Temperature of disinfectant solution should be minimum 20°C. The
minimum contact time is 5 minutes. If a different disinfectant is
used,
follow
consider the material compatibility for the medical device.
2) Immerge the Magnetic sensor attachment and spacer for EZU-RVS2 into
the disinfectant. (see figure 26). Set a clock to insure the
recommended contact time is 5 minutes.
3) Rinse the Magnetic sensor attachment and spacer for EZU-RVS2 with
running deionized water for approximately 1 minute.
4) Alternatively to step 3 suspend the Magnetic sensor attachment and
spacer for EZU-RVS2 in a tray filled with deionized water for 5
min.
5) Visually check the Magnetic sensor attachment and spacer for
EZU-RVS2 for leavings of the disinfectant. If necessary, repeat
the rinsing.
C) Biopsy attachment EZU-PA7B1
1) Before immersing the device, it is recommended to test the
concentration of disinfectant solution before each usage.
solution Cidex® OPA is ready for use and does not need to be diluted.
Test strips to verify that the appropriate concentration of Cidex®
OPA is correct are available by manufacturer. Test strips will
indicate a concentration above the Minimum Effective Concentration
(MEC). Please also note the expiration date of the test stripes.
Temperature of disinfectant solution should be minimum 20°C. The
minimum contact time is 5 minutes. If a different disinfectant is
used,
follow
consider the material compatibility for the medical device.
2) Immerge the biopsy attachment into the disinfectant. (see figure
27).
Set a clock to insure the recommended contact time is 5
minutes.
3) Rinse the biopsy attachment with running deionized water for 1
minute.
4) Alternatively to step 3 suspend the biopsy attachment in a tray
filled with deionized water for 5 min.
5) Visually
disinfectant. If necessary, repeat the rinsing.
the
manufacturer's
the
manufacturer's
check
the
biopsy
instructions.
instructions.
attachment
-29-
Please also
Please also
for
leavings
Q1E-EP1245
The
The
of
the