4
Operating procedures
Transition from exposure ready status to detector selection status
(Sleep)
Time to transition and transition conditions vary depending on the selected mode.
In Standard Synchronization Mode:
When the detector enters exposure ready status, the LED lamps on the ready
indicator light up or flash, and the ready indicator beeps. Capture an image within
10 minutes. If no operation is performed in exposure ready status for a maximum of
10 minutes (a maximum of 60 minutes* when connected to a Multi Box via a wiring
cable), exposure ready status is automatically changed to detector selection status
(Sleep). (If an operation is performed in exposure ready status, exposure ready
status is available for 10 minutes from that time.)
*The continuous operation time in exposure ready status is determined by the settings made at installation
time. For further details, consult your sales representatives.
In Non Generator Connection Mode:
The LED lamps on the ready indicator light up or flash and the indicator beeps to
indicates the beginning and end of exposure ready status of the detector,
countdown alarms, as well as X-ray detection
10 minutes. The detector's status remains exposure ready for a maximum of 10
minutes regardless of whether or not the user performs any operations during this
time. In 10 minutes, the detector enters detector selection status (Sleep)
automatically.
After exposure ready status ends, the detector enters detector selection status
(Sleep), and the READY LED lamp is no longer illuminated.
CAUTION
38
Information
Examination in Non Generator Connection Mode
• Avoid doing the following actions to the detector during
exposure ready status. Otherwise, the detector may acquire an
image without exposure.
– Giving a strong shock and vibration.
– Using the detector in a location where static electricity is easily
generated.
– Using the detector in a location where electromagnetic wave noise
might be generated.
• Images may not be acquired because the exposure conditions,
such as X-ray exposure conditions or target body positioning,
are not effective. For details, see "Troubleshooting" (➞ page 49).