3

Mounting the wet-section

Refer to Fig. 2 for wet-section dimensions. The wet-section weighs 4.5 kg (10 lb).
Note. Clearance – the enclosure doors can open 180°. If mounting in a confined area, allow sufficient clearance for door opening.
Dimensions in mm (in.)
386
(15.19)
Sample drain
(calibration)
Sample
outlet
1
*For M6 or
/
in. fastener.
4
Fig. 2 Low level dissolved oxygen wet-section dimensions
Referring to Fig. 3:
1. Mark the wall using the dimensions shown.
2. Drill and plug 3 holes
1
or
/
in. fixings.
4
A
3. Screw in top fixing
between the fixing head and the wall.
A
Note. Fixing
wet-section is placed over it.
4. Hang the wet-section onto fixing
firmly against the wall.
5. Secure the wet-section to the wall using 2 fixings
CI/ADS550–EN Rev. D
Ø 7 (0.27)*
480
(18.90)
210 (8.27)
210
REF
67
Sample inlet bulkhead connector –
(2.63)
suitable for either 6 mm or
A
B
and
in the wall suitable for M6
, leaving a gap of 20 mm (0.78 in.)
cannot
be
adjusted
once
A
, ensuring it is retained
B
281 (11.06)
281
DOOR OPEN
185 (7.28)
185
37 (1.45)
Sample drain (calibration)
1
/
in. pipe
4
Dimensions in mm (in.)
386
(15.19)
the
.
Fig. 3 Mounting the wet-section
Navigator 500
290 (11.41)
290
290
210
210
CTRS
CTRS
Sample inlet
bulkhead connector
Allow sufficient
210 (8.27)
clearance for
fully-open door
Ø 7
(0.27)*
80.2
(3.15)
5