GE BiliSoft 2.0 Operation Manual page 48

Phototherapy system
Table of Contents

Advertisement

Spectral Irradiance when
shipped (bare fiberoptic pad)
Using an Ohmeda Medical
BiliBlanket Meter II
Peak Wavelength
LED module estimated life
Sound level
X-ray
Light box (W x H x L)
Light box weight (excluding
fiberoptic pad)
Light Pad weight
Light Pad, size small
Light pad, large
Light Pad cable length
Service Life
IEC 60601-1; IEC 60601-1-11; IEC 60601-2-50; IEC 60601-1-2; IEC 60601-1-6
Note 1: The LED module life may vary when used in the actual clinical environment. Factors such as
duty cycle and ambient temperature may impact the life of the LED module.
7-2
Performance Specifications
Large Pad – 49
Small Pad – 70
Note: When the BiliSoft 2.0 fiberoptic light pad is inserted
into a BiliSoft Pad Cover or BiliSoft Nest, the average
spectral irradiance is 35 µW•cm
µW•cm
•nm
-2
445-470 nm
Under continuous use, at room temperature, a typical LED
1
module will run over 50,000 hours before the light intensity
drops 25% below minimum the rated output.
< 44 dB(A) at 1 meter
Light Pad is X-ray compatible
Physical Specifications
16.5 x 21 x 17.5 cm
< 1.7 kg
< .5 kg (large or small)
20 x 25 cm (light-emitting area)
25 x 30 cm (light-emitting area)
137 ± 5 cm
Product Life Span
The Light Box is designed to meet a life span of 7 years.
The Light Pad is designed for a life of 3.5 years. However,
with proper maintenance and repairs, the service life can
be extended as longs as service parts are available.
Regulatory Standards
ISO 10993-5; ISO 10993-10;
AAMI ES60601-1, CSA C22.2 No. 60601;
16CFR Part 1632.6 (for disposable pad covers)
µW•cm
•nm
(+/- 25%) 9-point check
-2
-1
µW•cm
•nm
(+/- 25%) 6-point check
-2
-1
•nm
-2
(small pad)
-1
(large pad) and 50
-1