Spectral Irradiance when
shipped (bare fiberoptic pad)
Using an Ohmeda Medical
BiliBlanket Meter II
Peak Wavelength
LED module estimated life
Sound level
X-ray
Light box (W x H x L)
Light box weight (excluding
fiberoptic pad)
Light Pad weight
Light Pad, size small
Light pad, large
Light Pad cable length
Service Life
IEC 60601-1; IEC 60601-1-11; IEC 60601-2-50; IEC 60601-1-2; IEC 60601-1-6
Note 1: The LED module life may vary when used in the actual clinical environment. Factors such as
duty cycle and ambient temperature may impact the life of the LED module.
7-2
Performance Specifications
Large Pad – 49
Small Pad – 70
Note: When the BiliSoft 2.0 fiberoptic light pad is inserted
into a BiliSoft Pad Cover or BiliSoft Nest, the average
spectral irradiance is 35 µW•cm
µW•cm
•nm
-2
445-470 nm
Under continuous use, at room temperature, a typical LED
1
module will run over 50,000 hours before the light intensity
drops 25% below minimum the rated output.
< 44 dB(A) at 1 meter
Light Pad is X-ray compatible
Physical Specifications
16.5 x 21 x 17.5 cm
< 1.7 kg
< .5 kg (large or small)
20 x 25 cm (light-emitting area)
25 x 30 cm (light-emitting area)
137 ± 5 cm
Product Life Span
The Light Box is designed to meet a life span of 7 years.
The Light Pad is designed for a life of 3.5 years. However,
with proper maintenance and repairs, the service life can
be extended as longs as service parts are available.
Regulatory Standards
ISO 10993-5; ISO 10993-10;
AAMI ES60601-1, CSA C22.2 No. 60601;
16CFR Part 1632.6 (for disposable pad covers)
µW•cm
•nm
(+/- 25%) 9-point check
-2
-1
µW•cm
•nm
(+/- 25%) 6-point check
-2
-1
•nm
-2
(small pad)
-1
(large pad) and 50
-1