Appendix B. Calibration and Testing with PanaView

B.1 Introduction
This appendix explains how to calibrate the Model XGM868i's analog outputs and inputs using PanaView™ interface
software. In addition, testing the optional totalizer, frequency and alarm relay outputs, and XGM hardware and
software is discussed. The appendix includes the following specific topics:
calibrating the
Slot 0
calibrating option card analog outputs
calibrating option card analog inputs and RTD inputs
testing option card alarm relays
testing option card totalizer and frequency outputs
uploading XGM868i signal array data and memory data
testing the EPROM, RAM and DAC
setting the number of channels
resetting the XGM868i to default values
The Model XGM868i electronics enclosure includes one slot for installation of an input/output option card. This slot is
designated as
, and a variety of option cards are available. In addition, every Model XGM868i flowmeter includes
Slot 1
two built-in analog outputs (A and B) at terminal block
Note: Convention in this manual identifies any expansion slot as
only for data logging or RS485 option cards.
See Chapter 1, Installation, and Appendix B, Data Records, in the Startup Guide for a complete description of the
available option cards and the procedures for wiring them.
DigitalFlow™ XGM868i Service Manual
analog outputs
Appendix B. Calibration and Testing with PanaView
, which is designated as
J1
, where
Slot X
.
Slot 0
is a number from 0-2.
X
is used
Slot 2
41