By examination of an early pregnancy the exposure time shall be as
short as possible. Start examination with acoustic output power set
to L (Low).
The probe connector is not waterproof. Do not allow liquid to contact
the connector.
Do not drop, hit or bend the probe.
Use only water, detergents and disinfectants in the suppliers list.
Under sterile condition, use appropriate protection for probe and
cable. Some Latex material may create allergic reactions.
The probe and the attachment are not delivered disinfected or
sterilized.
The acoustic lens of the probe is manufactured very thin and delicate
to get the high resolution.
Therefore, in case of wiping out the ultrasound jelly or cleaning
the surface of the acoustic lens, please use the soft cloth or tissue
paper and handle with care.
Do not use this probe with other equipment except for those
specifically approved in the manual. Use with unapproved equipment
can result in an electric shock, burn, or other injury to the patient
or operator and damage to the probe and the other equipment.
Do not wash, disinfect or sterilize using procedures other than those
specified in this manual. Infection could result due to incomplete
washing disinfection or sterilization. It can also result in damage
to the probe or reduced performance.
CAUTION
-37-
Q1E-EP1457