Installing The Holder With Changeover Mechanism - Miele A 865 Operating Instructions Manual

For reprocessing canulae from 3f and mf syringes
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 25
Installing the
holder with
changeover
mechanism
Depending on the geometry of the 3F and MF syringes, the holder 
(highlighted in grey) can be installed rotated through 180°
(changeover mechanism).
Place the compression spring  inside the sleeve .
Push the holder  onto the compression spring  and fit the lid .
Screw the hex nut  onto the sleeve  as far as the stop.
Connect the spring clip  to the thread of the sleeve .
Holder facing downwards (changeover mechanism)
a
Lid
b
Holder
c
O-ring
d
Compression spring
e
Sleeve
f
Spring clip
g
Hex nut
en - Areas of application
31