This is the setting instructions for the Rel. Hum. and Pressure dew point and Temp.
12.3.2.3.3 Modbus settings for the METPOINT
When connecting the METPOINT
required:
First step: select a free sensor channel
Main menu Settings Sensor settings Select a free channel (example: channel A1)
Second step: select Modbus type
Main menu Settings Sensor settings A1 Type text field Select Modbus and confirm with
>OK<.
Third step: define a name
Main menu Settings Sensor settings A1 Name text field
Now, a name needs to be entered.
Fourth step: define the Modbus settings
Main menu Settings Sensor settings A1 Modbus settings
®
METPOINT
BDL portable
SD23
®
®
SD23 via Modbus, the following settings are
The corresponding Modbus ID can be taken
from the data sheet of the sensor (here, for
example, 1).
Adjust the other parameters according to the
illustration.
Operation METPOINT® BDL portable