The Tray position indicator will display as Upper, Lower, NP lens for plates, and None.
SNOW
™
mode
The Signal to Noise Optimization Watch (SNOW) reduces noise and avoids saturation of
the image. Multiple exposures are captured and merged by averaging the results into
a single image. This improves the signal-to-noise ratio by removing artefacts and reducing
false positive signals. This setting can be used for Chemiluminescence and Fluorescence
imaging methods.
Amersham ImageQuant 800 User Manual 29383813 AA
3 Handle methods in ImageQuant 800 control software
3.2 General settings
39