Chapter 2 Remote Interface Operation

Phase-Lock Error Messages

Phase-Lock Error Messages

This section lists the two error messages that can be generated if the
Phase-Lock option is installed. Refer to chapter 5 in the HP 33120A
User's Guide for a complete listing of error messages.
-221
Settings conflict; cannot adjust phase in present configuration
Option 001 Phase-Lock Only. The phase cannot be adjusted real-time if
an arbitrary waveform is selected, a modulation mode (other that burst)
is enabled, or if burst is enabled with a burst count other than infinity.
580
Phase-locked loop is unlocked
Option 001 Phase-Lock Only. The function generator has detected an
"unlock" condition. You must execute the PHAS:UNL:ERR:STAT ON
command to enable this error.