Replacing Casters On The Mac 1600 Trolley - GE Modular MAC Trolley Service Manual

Ecg trolley
Table of Contents

Advertisement

3.
Remove the cover from the base.
See
4.
Remove the hex nut from the caster to be replaced.
Discard the caster. If the hex nut is undamaged, it can be retained to use on
the new caster.
5.
Repeat this procedure for the other casters.
6.
Reverse the disassembly procedures to reassembly the trolley.
7.
Reattach the MAC device and trolley options.

Replacing Casters on the MAC 1600 Trolley

This procedure applies only to the MAC 1600 Trolley. For instructions on removing the
casters from the Modular MAC Trolley, refer to
MAC Trolley" on page
Before removing the base, remove the MAC unit and accessories from the trolley. See
"Removing the MAC Unit from the MAC 1600 Trolley" on page 23
replacing the MAC unit to the MAC 1600 Trolley.
2044854-012E
"Replacing the Base Cover " on page 28
28.
MAC™ ECG Trolley
for instructions.
"Replacing Casters on the Modular
for instructions on
Maintenance
29