Electromagnetic Compatibility (Emc) - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Priming the circu-
lation pump
Protective foil/
countertop pro-
tector
 The gaps between a built-in machine and adjacent cabinetry
must not be sealed, e.g., with silicone sealant, as this could com-
promise the priming of the circulation pump.
The protective foil supplied protects the countertop from damage
caused by steam when the door is opened. It should be positioned
underneath the countertop above the machine door.

Electromagnetic compatibility (EMC)

The machine has been tested for electromagnetic compatibility (EMC)
in accordance with EN 61326-1 and is suitable for operation in com-
mercial environments, such as hospitals, medical practices, and labo-
ratories plus other similar environments which are connected to the
domestic electrical supply.
The machine's high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic machines in the
vicinity.
Flooring in the installation area must be wood, concrete, or tiled. Syn-
thetic flooring must be able to withstand a relative humidity level of at
least 30% to minimize the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment. Check that the power
supply voltage is within a range of +/-10% of its nominal value.