Accelerometer Interface Module Operation Manual
5.
Unscrew the four
screws at each
corner of the lid.
6.
7.
Adjust the potentiometer until the multimeter reads 0.3535 ±7.0 Vrms.
Close and tighten the lid and verify that the multimeter still reads 0.3535 Vrms. If it
does not then repeat steps 6 and 7. Note that when the lid is open the circuitry is
unshielded and the charge amp can pick up noise.
20
Open the lid and fold it back.