Cordless fetal transducer system
Hide thumbs Also See for Avalon CTS M2720A:

Advertisement

M2720A Avalon CTS
Cordless Fetal Transducer System
Quick Start Guide
110 - 240 V AC
Remove transducer
The transducers are pre-warmed up to body temperature in the base station. Please advise your
patient before applying the transducer.
Changing from US to DECG monitoring
1
Dock the US transducer.
2
Take out the ECG transducer when the lamp
slot is green.
3
Connect the DECG adapter cable M1362B and the fetal scalp
electrode to the ECG transducer.
4
Start DECG monitoring.
Two active transducers transmitting FHR to the same base station cause error message E9. To clear, dock one transducer.
After Use: Clean transducers. Then dock in appropriate base station slot.