5 Shutting down / removing the sensor
assembly from the process
DANGER – Serious damage to health / risk to life
Allow sufficient time for the sensor assembly to cool
before performing these procedures.
5.1 Shutting the sensor assembly down at the process
Referring to Fig. 5.1.
1. Close the air supply valve A and shut down the test gas
line B at the supply.
Fig. 5.1 Shutting down instrument air and test gas supplies
5.2 Disconnecting the sensor air and test gas supplies at
the process
Referring to Fig. 5.2:
1. Disconnect air line B and test gas line C at sensor D.
Fig. 5.2 Disconnecting instrument air and test gas supplies
5.3 Disconnecting the sensor electrical power and
signal cables at the process
DANGER – Serious damage to health / risk to life
Allow sufficient time for the sensor assembly to cool
before performing this procedure.
DANGER – Serious damage to health / risk to life
The transmitter must be isolated from mains power
supplies before performing this procedure.
Referring to Fig. 5.3:
1. Use a medium flat-bladed screwdriver to unscrew 4
(captive) cover screws A and remove cover B from
sensor assembly C.
2. Disconnect mains cable D from terminal block E
3. Disconnect signal cable F from terminal block G.
4. Disconnect thermocouple cables H from terminal block
I.
5. If optional blowback is fitted, disconnect cable J from
terminal block K.
Fig. 5.3 Disconnecting sensor cables