Maintenance
7.2.2 Entering the battery capacity
The battery capacity must be re-entered after changing a
battery.
If the device is operated with high-level security (see appli-
cation manual) then the battery capacity can only be en-
tered using the enSuite software!
◼
To open the administrator lock move the cursor to the "Admin" register
and then via the following path to the "Cod.A" value
(Administrator code entry):
Admin. Device settings Access Cod.A
After delivery, the administrator code for keyboard input is
"00000000".
Press the ENTER button. One of the value's digits flashes.
◼
Move the cursor to the digits using the or arrow keys and update
◼
them using the or arrow keys.
◼
Once you have changed all of the digits, press the ENTER key in order
to confirm your entry. The input value can be deleted by pressing the
ESC button.
◼
To enter the battery capacity move the cursor to the "Serv." register and
to the "Bat.K" value (battery capacity) via the following path:
Admin Batteries Bat.C
Press the ENTER button. One of the value's digits flashes.
◼
Move the cursor to the digits using the or arrow keys and update
◼
them using the or arrow keys.
When using one size D battery, the value 13.0 Ah should
be entered for "Bat.C". When using 2 batteries, the value
26.0 Ah must be entered.
◼
Once you have changed all of the digits, press the ENTER key in order
to confirm your entry. The input value can be deleted by pressing the
ESC button.
◼
To close the administrator lock move the cursor to the "Admin" register
and to the "St.AS" value (Status of administration lock) via the following
path:
Admin. Device settings Access St.AL
Press the ENTER key "open" flashes.
◼
79