Dimensions And Mass; Siting Requirements - Toshiba Activion 16 Product Data

Multislice helical ct scanner
Table of Contents

Advertisement

Image storage and archiving
• The system is provided with a 217 Gbyte magnetic hard
disk as standard equipment, permitting the on-line stor-
age of approximately 100,000 images and 3,600 rota-
tions of raw data.
• A 9.4 Gbyte DVD-RAM disk drive is provided as stan-
dard equipment. The image storage capacity of the
DVD-RAM disk is approximately 16,000 images per disk.
Image filming
• Filming of images can be performed manually or auto-
matically from the console.
• Automatic filming sends an entire study to the laser cam-
era. Filming is performed in background mode so that
other scanner and image processing functions can be
performed without interruption or delay.
• When T-mode is used, related information items dis-
played together with an image (surrounding the image, in
a small font) are displayed in the footer area using a larg-
er font, permitting not only easier reading but also sim-
pler film management.
Note: To use T-mode, the laser imager must support
2048 pixels x 2404 pixels for a 1 x 1 frame.
Patient throughput
Patient throughput and cost effectiveness were major
objectives in the design and production of the system.
• The system incorporates a 4.0-MHU X-ray tube with a
very fast cooling rate of 864 kHU/min in actual use.
• High-speed scans can be performed in as little as 0.75
second per scan.
• Real-time reconstruction is possible in scanoscopy.
• CT images can be reconstructed in 0.5 second for 0.75-
second routine scans.
• Ease of operation is ensured by incorporating a hybrid
keyboard, mouse-driven menus, and a large color LCD
screen.
• The couch-top can be lowered very near the floor, simpli-
fying patient transfer.
COMPLIANCE
Council Directive 93/42/EEC
Concerning Medical Devices (Medical Device Directive)
• IEC:
IEC 60601-1: 1988
IEC 60601-1 Amd.1: 1991
IEC 60601-1 Amd.2: 1995
IEC 60601-1-1: 2000
IEC 60601-1-2: 2001
IEC 60601-1-3: 1994
IEC 60601-1-4 Ed.1.1: 2000
IEC 60601-2-28: 1993
IEC 60601-2-32: 1994
IEC 60601-2-44: Ed.2.1: 2002

DIMENSIONS AND MASS

Dimensions L × W × H
Unit
mm (in)
900 × 2,070 × 1,910
Gantry
(35.4 × 81.5 × 75.2)
2,690 × 630 × 450
Long patient
(105.9 × 24.8 × 17.7)
couch version
Patient
2,390 × 630 × 450
couch
Short patient
(94.1 × 24.8 × 17.7)
couch version
815 × 450 × 700
CPU cabinet 1
(32.1 × 17.7 × 27.6)
815 × 450 × 700
CPU cabinet 2
(32.1 × 17.7 × 27.6)

SITING REQUIREMENTS

Power requirements
• Phase:
Three-phase
• Voltage:
200 V*
• Frequency:
50 Hz or 60 Hz ±0.5 Hz
• Line capacity:
75 kVA
• Voltage fluctuation
due to load variation:
Less than 5%
• Power voltage
fluctuation:
Less than 10%**
* Please consult Toshiba in the case of other voltages or exces-
sive power fluctuation.
** Represents the total voltage fluctuation due to load and power
variation.
Grounding
Grounding must be provided in accordance with local
regulations for medically used electrical equipment.
MPDCT0290EA
Mass
kg (lb)
1,280
(2,822)
450
(992)
420
(926)
100
(220)
85
(187)
9