General Specifications; Block Diagram - Siemens SITRANS RD500 Operating Instructions Manual

Remote data manager module 8 inputs, 6 relay outputs
Hide thumbs Also See for SITRANS RD500:
Table of Contents

Advertisement

GENERAL SPECIFICATIONS

1. POWER: Derived from system backplane. (The module draws 170 mA max. load on
power input of RD500). Modules may be hot-swapped (replaced while powered
up).
2. LEDs:
STS - Status LED shows module condition.
IN1-IN8 - LEDs are lit when associated input is active.
ALM - Alarm LED is lit when an internal alarm condition exists.
3. MEMORY: Non-volatile memory retains all programmable parameters. RD500
also stores the parameters in order to reprogram modules that are replaced.
4. INPUTS: DIP switch selectable for sink or source
Maximum voltage: +30 VDC, reverse polarity protected
Off Voltage: < 1.2 Volts
On Voltage: > 3.8 Volts
Input Impedance: Source Mode 10K ohms; Sink Mode 20K ohms
Input Frequency*:
Filter switch on: 50 Hz
Filter switch off: 300 Hz
* Actual useable frequency limited by communication to external device.
5. OUTPUTS:
Form A Relay Output:
Type: N.O.
The following pairs of relays share the common terminal: 1&2, 3&4, 5&6
Current Rating by pair: 3 Amps @ 30 VDC / 125 VAC resistive
1/10 HP @ 125 VAC
Life Expectancy: 200,000 cycles at maximum load rating. (Decreasing load,
increasing cycle time, and use of surge suppression such as RC snubbers
increases life expectancy.)
6. ISOLATION LEVEL: 500 Vrms @ 50/60 Hz for 1 minute between the following:
Inputs
Outputs
RD500 Power Supply Input
7. COMMUNICATIONS: Provided by the RD500
8. ENVIRONMENTAL CONDITIONS:
Operating Temperature Range: 0 to +50°C
Storage Temperature Range: -40 to +85°C
Operating and Storage Humidity: 85% max relative humidity, non-condensing,
from 0 to +50°C
Vibration According to IEC 68-2-6: Operational 10 to 150 Hz, 0.075 mm
amplitude in X, Y, Z direction 1 g.
Shock According to IEC 68-2-27: Operational 25 g's (10 g's relay), 11 msec in 3
directions.
Altitude: Up to 2000 meters
+
24VDC
7ML19985MD01

BLOCK DIAGRAM

PORT 1
PROGRAMMING
A
PORT 2
COMMUNICATIONS
A
ISOLATED
PORT 3
ETHERNET
B
POWER
-
SUPPLY
A
RD500
SITRANS RD500
9. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL508, CSA C22.2 No. 14-M05, UL61010-1, CAN/CSA C22.2 No. 61010-1
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment for
measurement, control, and laboratory use, Part 1.
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: 2006: Electrical Equipment for
Measurement, Control and Laboratory use.
Immunity to Industrial Locations:
Electrostatic discharge
Electromagnetic RF fields
Fast transients (burst)
I/O signal connectected to power
Surge
RF conducted interference
Emissions:
Emissions
Notes:
1. Criterion A: Normal operation within specified limits.
2. Criterion B: Temporary loss of performance from which the unit
self-recovers.
3. Power supplied from back plane via RD500 Module.
10. CONSTRUCTION: Case body is high impact plastic. For indoor use only.
Installation Category II, Pollution Degree 2.
11. CONNECTIONS: Removable wire clamp screw terminal blocks.
Wire Gage: 28-16 AWG terminal gage wire
Torque: 1.96-2.23 inch/lbs (0.22-0.25 N-m)
12. MOUNTING: Snaps on to standard DIN style top hat (T) profile mounting
rails according to EN50022 -35 x 7.5 and -35 x 15.
13. WEIGHT: 6.6 oz (187.1 g)
ISOLATED
ISOLATED
1
INPUTS
8
RELAY
OUTPUTS
E
F
POWER
SUPPLY
Module
EN 61000-4-2
Criterion B
4 kV contact discharge
8 kV air discharge
EN 61000-4-3
Criterion A
10 V/m
EN 61000-4-4
Criterion B
power
2 kV
I/O signal
1 kV
2 kV
EN 61000-4-5
Criterion B
power
1 kV L-L, 2 kV L-G
signal
1 kV
EN 61000-4-6
Criterion A
3 V/rms
EN 55011
Class A
1
COM
1,2
2
3
COM 3,4
4
5
COM 5,6
6
Page 2

Advertisement

Table of Contents
loading

Table of Contents