Measurement and Analysis
7. To end the measurement, press the middle trackball button.

Measuring M-Mode Distance

1. Start M-mode by pressing the M-mode soft key, and then press the middle trackball button
to update the M-mode display.
2. Obtain the desired M-mode display and press Freeze.
3. Press or touch Measure, and then touch Tools.
4. Touch Distance.
5. Use the trackball to position the vertical time caliper and press the left or right trackball
button.
6. Use the trackball to position the horizontal depth caliper. The distance between the depth
calipers appears in the results.
7. To make additional measurements, repeat step 5 and step 6.
8. Press the middle trackball button to complete the measurement.

Measuring the Doppler Trace on a Sweeping Display

The measurements that appear depend on the settings for Doppler in the Tools and Results
setups.
You can measure Doppler velocity on a live image.
1. Press PW.
2. Press the middle trackball button to update the Doppler display.
3. Obtain the Doppler image you want to measure (spectral waveform).
4. Touch Measure.
5. Use the trackball to position the horizontal cursor.
6. There are no measurement values in the results; you can read the velocity or frequency
value from the velocity or frequency scale.

Obtaining a Typical Labeled Measurement

This general procedure describes how to measure by using a typical labeled measurement tool.
1. Obtain the image you want to measure and press Freeze.
2. Press or touch Measure.
3. Do one or more of the following:
EPIQ 7 User Manual 4535 617 25341
On the Calc Package tab of the touch screen, touch a collection and then a
measurement label.
Performing an Exam
169